These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Performance characteristics of four automated natriuretic peptide assays. Author: Rawlins ML, Owen WE, Roberts WL. Journal: Am J Clin Pathol; 2005 Mar; 123(3):439-45. PubMed ID: 15716241. Abstract: Measurement of circulating B-type natriuretic peptide (BNP) and N-terminal proBNP (NT-proBNP) can identify patients with heart failure and guide therapy. The limit of detection, linearity, imprecision, method comparison, analytic concordance, and reference intervals of the Access 2 BNP (Biosite, San Diego, CA), ADVIA Centaur BNP (Bayer Diagnostics, Tarrytown, NY), AxSYM BNP (Abbott Diagnostics, Abbott Park, IL), and E170 NT-proBNP (Roche Diagnostics, Indianapolis, IN) methods were evaluated. The Triage meter BNP assay (Biosite) was the comparison method. Imprecision testing showed total coefficients of variation of 4.1%, 4.4%, 5.5%, and 0.8% for the Access 2, ADVIA Centaur, AxSYM, and E170, respectively. Relative to the Triage meter, method comparison revealed a slope of 0.96 and r = 0.95, a slope of 0.77 and r = 0.92, a slope of 1.13 and r = 0.94, and a slope of 8.8 and r = 0.80 for the Access 2, ADVIA Centaur, AxSYM, and E170, respectively. Overall analytic concordance values with the Triage meter were 95.9%, 92.9%, 92.4%, and 84.3% for the Access 2, ADVIA Centaur, AxSYM, and E170, respectively. All automated natriuretic peptide methods showed acceptable analytic performance.[Abstract] [Full Text] [Related] [New Search]