These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Electron beam dosimetry of Total Skin Electron Therapy (TSET). Author: Piermattei A, Rossi G, Azario L, Fidanzio A, Balducci M, Valentini V. Journal: Radiol Med; 2004; 108(5-6):549-59. PubMed ID: 15723001. Abstract: Total Skin Electron Therapy (TSET) was carried out using an electron beam with a nominal energy of 6 MeV. The beam was adequately filtered and angled in order to create dual fields. The uniformity of the dose distribution to the patient was 10%. The relative dosimetry of the dual beam was performed using a silicon diode and an ionization chamber in a standard water phantom. X-Omat V films were irradiated in a cylindrical PMMA phantom in order to obtain the dose distribution for the six TSET dual fields used in the treatment. Absolute dosimetry was carried out with a calibrated ionization chamber placed in a cylindrical water phantom. The dose contribution per monitor unit of the single dual beams was determined with this method.[Abstract] [Full Text] [Related] [New Search]