  • Title: Noise susceptibility of cochlear implant users: the role of spectral resolution and smearing.
    Author: Fu QJ, Nogaki G.
    Journal: J Assoc Res Otolaryngol; 2005 Mar; 6(1):19-27. PubMed ID: 15735937.
    Abstract:
    The latest-generation cochlear implant devices provide many deaf patients with good speech recognition in quiet listening conditions. However, speech recognition deteriorates rapidly as the level of background noise increases. Previous studies have shown that, for cochlear implant users, the absence of fine spectro-temporal cues may contribute to poorer performance in noise, especially when the noise is dynamic (e.g., competing speaker or modulated noise). Here we report on sentence recognition by cochlear implant users and by normal-hearing subjects listening to an acoustic simulation of a cochlear implant, in the presence of steady or square-wave modulated speech-shaped noise. Implant users were tested using their everyday, clinically assigned speech processors. In the acoustic simulation, normal-hearing listeners were tested for different degrees of spectral resolution (16, eight, or four channels) and spectral smearing (carrier filter slopes of -24 or -6 dB/octave). For modulated noise, normal-hearing listeners experienced significant release from masking when the original, unprocessed speech was presented (which preserved the spectro-temporal fine structure), while cochlear implant users experienced no release from masking. As the spectral resolution was reduced, normal-hearing listeners' release from masking gradually diminished. Release from masking was further reduced as the degree of spectral smearing increased. Interestingly, the mean speech recognition thresholds of implant users were very close to those of normal-hearing subjects listening to four-channel spectrally smeared noise-band speech. Also, the best cochlear implant listeners performed like normal-hearing subjects listening to eight- to 16-channel spectrally smeared noise-band speech. These findings suggest that implant users' susceptibility to noise may be caused by the reduced spectral resolution and the high degree of spectral smearing associated with channel interaction. Efforts to improve the effective number of spectral channels as well as reduce channel interactions may improve implant performance in noise, especially for temporally modulated noise.
    [Abstract] [Full Text] [Related] [New Search]