These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Perchlorate-selective membrane electrode based on a new complex of uranil.
    Author: Mazloum Ardakani M, Jalayer M, Naeimi H, Zare HR, Moradi L.
    Journal: Anal Bioanal Chem; 2005 Mar; 381(6):1186-92. PubMed ID: 15770475.
    Abstract:
    A potentiometric ion-selective electrode based on new compound, as a carrier, has been successfully developed for detection of perchlorate anion in aqueous solution. Within the perchlorate ion concentration range 1.0x10(-6) to 1.0 mol L(-1) the electrode had a linear response with a Nernstian slope of 60.6+/-1.0 mV per decade . The limit of detection as determined from the intersection of the extrapolated linear segments of the calibration plot was 8.0x10(-7) mol L(-1). The proposed electrode has fairly a good discriminating ability towards ClO(4) (-) ion in comparison to other anions. The sensor has a response time of < or =10 s and can be used for at least 2 months without substantial divergence in potential. It was successfully applied to direct determination of perchlorate in urine and water.
    [Abstract] [Full Text] [Related] [New Search]