  • Title: Radiation exposure of patients undergoing whole-body dual-modality 18F-FDG PET/CT examinations.
    Author: Brix G, Lechel U, Glatting G, Ziegler SI, Münzing W, Müller SP, Beyer T.
    Journal: J Nucl Med; 2005 Apr; 46(4):608-13. PubMed ID: 15809483.
    Abstract:
    UNLABELLED: We investigated radiation exposure of patients undergoing whole-body 18F-FDG PET/CT examinations at 4 hospitals equipped with different tomographs. METHODS: Patient doses were estimated by using established dose coefficients for 18F-FDG and from thermoluminescent measurements performed on an anthropomorphic whole-body phantom. RESULTS: The most relevant difference between the protocols examined was the incorporation of CT as part of the combined PET/CT examination: Separate low-dose CT scans were acquired at 2 hospitals for attenuation correction of emission data in addition to a contrast-enhanced CT scan for diagnostic evaluation, whereas, at the other sites, contrast-enhanced CT scans were used for both purposes. Nevertheless, the effective dose per PET/CT examination was similar, about 25 mSv. CONCLUSION: The dosimetric concepts presented in this study provide a valuable tool for the optimization of whole-body 18F-FDG PET/CT protocols. Further reduction of patient exposure can be achieved by modifications to the existing hardware and software of PET/CT systems.