These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Study on RP-HPLC fingerprint of compound hongcao injection].
    Author: Lan YY, Liu LN, Wang YL, Li YJ, Huang Y, He X, Zheng L, Wang AM.
    Journal: Zhongguo Zhong Yao Za Zhi; 2005 Mar; 30(6):440-2. PubMed ID: 15810449.
    Abstract:
    OBJECTIVE: To establish the fingerprint detecting standard of Compound Hongcao Injection by reversed phase high performance liquid chromatography. METHOD: The chromatographic conditions were as follows: an Inersil-ODS-3 column was used; the mobile phases (acetontrile and 0.1% phosphoric acid) with gradient elution; the flow rate 1.0 mL x min(-1) and the UV absorbance detection at 300 nm. RESULT: Under the same selected chromatographic conditions, the HPLC fingerprints of raw drugs and their intermediate, finished products were obtained with good separation and correlation according to the technical requirements of fingerprint on Injection of Chinese traditional medicine. CONCLUSION: 17 common peaks as well as their retention times and peak area ratios on HPLC fingerprint can be used as the important parameters of the quality control for Compound Hongcao Injection.