These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Comparative study of quantitative blood pool SPECT imaging with 180 degrees and 360 degrees acquisition orbits on accuracy of cardiac function. Author: Adachi I, Umeda T, Shimomura H, Suwa M, Komori T, Ogura Y, Utsunomiya K, Kitaura Y, Narabayashi I. Journal: J Nucl Cardiol; 2005; 12(2):186-94. PubMed ID: 15812373. Abstract: BACKGROUND: Quantitative blood pool single photon emission computed tomography (SPECT) (QBS) can measure ejection fraction (EF) and volumes from gated blood pool single photon emission tomography (GBPS) working in fully automatic mode in 3-dimensional space. The effects of 180 degrees and 360 degrees data acquisition in GBPS have not been fully evaluated. This study compares the accuracy of 360 degrees and 180 degrees data acquisition for left ventricular (LV) systolic function in a clinical study and measures LV volume by GBPS compared with ultrasound echocardiography. METHODS AND RESULTS: The study population comprised 9 normal volunteers and 34 patients. GBPS data were acquired by use of 360 degrees rotation and 60 stops per head. All 60 (360 degrees ) and 30 (45 degrees right anterior oblique to 45 degrees left posterior oblique) pieces of projection data that were selected for reconstructing the 180 degrees data were reconstructed and both ventricular functional parameters were automatically obtained by QBS software. The contour of the LV septal wall was concave in 6 patients (14%) when processed at 180 degrees , whereas a concave septum at 360 degrees processing was observed in only 1 patient (2%). The coefficients of correlation between 180 degrees and 360 degrees were 0.467 for the end-diastolic volume (EDV) and 0.648 for the end-systolic volume (ESV). The mean 180 degrees EDV value (152.9 +/- 46.1 mL) was significantly smaller than that of the 360 degrees EDV (191 +/- 70.8 mL) ( P < .001). However, there was no significant difference between the 360 degrees EDV (0.623) and 180 degrees EDV (0.407) as compared by echocardiography ( P = .218). The agreement of the EF between both methods was close ( r = 0.894, P < .0001). The agreement of the right ventricular volumes between the 180 degrees and 360 degrees orbits was close ( r = 0.800 for EDV and 0.706 for ESV). The EF was relatively dispersed between the 180 degrees and 360 degrees methods ( r = 0.642). CONCLUSION: This study showed that SPECT image acquisition by use of both the 180 degrees method and the 360 degrees method considerably underestimated LV volume quantification. In addition, the LV volume with the 180 degrees method was significantly smaller than that with the 360 degrees method. Thus a 360 degrees acquisition orbit may be suitable for more quantitatively accurate results when blood pool imaging is performed with gated SPECT.[Abstract] [Full Text] [Related] [New Search]