  • Title: Comparison of dosimetric standards of USA and France for HDR brachytherapy.
    Author: Douysset G, Gouriou J, Delaunay F, DeWerd L, Stump K, Micka J, Laboratoire National Henri Becquerel, University of Wisconsin Accredited Dosimetry Calibration Laboratory.
    Journal: Phys Med Biol; 2005 May 07; 50(9):1961-78. PubMed ID: 15843730.
    Abstract:
    A bilateral comparison of national dosimetric standards for high dose rate brachytherapy has been conducted between the Laboratoire National Henri Becquerel and the University of Wisconsin Accredited Dosimetry Calibration Laboratory. A complete overview of the methods that are currently in use to establish the two national standards is given. The comparison has been carried out using well-type transfer ionization chambers. Three units have been calibrated in both laboratories, and calibration coefficients have been compared. The discrepancies between the two measurements are within 0.3%.