These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Comparison of dosimetric standards of USA and France for HDR brachytherapy. Author: Douysset G, Gouriou J, Delaunay F, DeWerd L, Stump K, Micka J, Laboratoire National Henri Becquerel, University of Wisconsin Accredited Dosimetry Calibration Laboratory. Journal: Phys Med Biol; 2005 May 07; 50(9):1961-78. PubMed ID: 15843730. Abstract: A bilateral comparison of national dosimetric standards for high dose rate brachytherapy has been conducted between the Laboratoire National Henri Becquerel and the University of Wisconsin Accredited Dosimetry Calibration Laboratory. A complete overview of the methods that are currently in use to establish the two national standards is given. The comparison has been carried out using well-type transfer ionization chambers. Three units have been calibrated in both laboratories, and calibration coefficients have been compared. The discrepancies between the two measurements are within 0.3%.[Abstract] [Full Text] [Related] [New Search]