Title: A double-line sequential injection system for the spectrophotometric determination of copper, iron, manganese, and zinc in waters. Author: Morais IP, Souto MR, Rangel AO. Journal: J AOAC Int; 2005; 88(2):639-44. PubMed ID: 15859092. Abstract: A double-line sequential injection system was developed for the spectrophotometric determination of several metal ions in waters. The proposed double-line configuration was used to enable adding sample and chromogenic reagents as merging zones. The methodology was applied to the spectrophotometric determination of copper, iron, manganese, and zinc in samples of diverse origins at the range of 0.15-5.00, 0.10-10.0, 0.48-4.00, and 0.11-5.00 mg/L, respectively. Different chromogenic reagents and detection wavelengths were used. The chromogenic reagents for iron and manganese were 1,10-phenanthroline and formaldoxime, respectively. Copper and zinc were both determined using the analytical reagent zincon. Analytical characteristics of the methodology, such as manifold parameters, buffer pH, and reagent concentrations were optimized, and interference of some of the metal ions commonly present in water sample was assessed. Results of the analysis were in agreement with those obtained by atomic absorption spectrometry. Repeatability, expressed as the relative standard deviation for 10 consecutive injections of water samples, was lower than 6%. The determination rate was approximately 36/h.[Abstract] [Full Text] [Related] [New Search]