  • Title: High performance liquid chromatographic methods for the determination of aripiprazole with ultraviolet detection in rat plasma and brain: application to the pharmacokinetic study.
    Author: Shimokawa Y, Akiyama H, Kashiyama E, Koga T, Miyamoto G.
    Journal: J Chromatogr B Analyt Technol Biomed Life Sci; 2005 Jul 05; 821(1):8-14. PubMed ID: 15897016.
    Abstract:
    High performance liquid chromatographic (HPLC) methods were validated for the determination of aripiprazole (OPC-14597, Abilify) in rat plasma and brain. Separation was by Nova-pak phenyl column; flow rate, 1.0 ml/min; mobile phase, acetonitrile-methanol-20 mM sodium sulfate-acetic acid (27:25:48:1, v/v/v/v); UV detection at 254 nm. Reproducibility in plasma and brain showed excellent precision (within 7.8 and 10.6%) and accuracy (96.0-102.4% and 99.0-108.7%) with calibration curve ranges 10.0-2000 ng/ml and 30.0-6000 ng/g, respectively. Validated HPLC methods were successfully applied to pharmacokinetic study of aripiprazole in rats, demonstrating brain concentrations after oral administration five times higher than plasma concentrations.