  • Title: Flow injection analysis of riboflavin with chemiluminescence detection using a N-halo compounds-luminol system.
    Author: Safavi A, Karimi MA, Nezhad MR.
    Journal: Luminescence; 2005; 20(3):170-5. PubMed ID: 15924325.
    Abstract:
    A chemiluminescent method for the determination of riboflavin is described. The method is based on the chemiluminescence (CL) generated during the oxidation of luminol by N-bromosuccinimide (NBS) and N-chlorosuccinimide (NCS) in alkaline medium. It was found that riboflavin could greatly enhance this CL intensity when present in the luminol solution. Based on this observation, a new flow-injection CL method for the determination of riboflavin is proposed in this paper. The detection limits were 7.5 ng/mL and 3.5 ng/mL of riboflavin for the NBS- and NCS-luminol CL systems, respectively. The relative CL intensity was linear, with riboflavin concentration in the range 19-600 ng/mL and 600-2000 ng/mL for the NBS-luminol CL system, and 12-200 ng/mL and 200-2000 ng/mL for the NCS-luminol CL system. The results obtained for the assay of pharmaceutical preparations compared well with those obtained by the official method and demonstrated good accuracy and precision.