These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Fiber-optic transmission catheter for regional venous oxygen saturation and blood flow. Author: Iwasaki H, Yoshizaki K, Koyano H. Journal: J Appl Physiol (1985); 1992 Apr; 72(4):1616-21. PubMed ID: 1592756. Abstract: We have developed a method for monitoring regional venous oxygen saturation. The key feature of this system is the use of highly flexible polymer fiber optics, and this flexibility allowed the production of a new fiber-optic transmission catheter. The space between the "face-to-face" positioned fiber-optic tips forms a remote catheter-based transmission cell. Our method applies Twersky's theory, in which absorption and scattering can be treated independently. Fresh rabbit blood was pumped through a disk oxygenator in which gas exchange occurred and passed the catheter. Simultaneous results obtained by the catheter and a cuvette oximeter were excellent (r = 0.99, SD = 1.1%). Oxygen saturation measured by this catheter was independent of vessel wall artifacts, blood pH, and flow velocity. Another application of this method is measurement of blood flow by the dye- (indocyanine green) dilution technique. The results of flow measurements by the catheter appeared to be satisfactory (r = 0.99, SD = 1.7%). This study concludes that our method is effective for monitoring the balance between regional oxygen supply and demand.[Abstract] [Full Text] [Related] [New Search]