These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Measurement of the modulation transfer function of digital X-ray detectors with an opaque edge-test device.
    Author: Illers H, Buhr E, Günther-Kohfahl S, Neitzel U.
    Journal: Radiat Prot Dosimetry; 2005; 114(1-3):214-9. PubMed ID: 15933111.
    Abstract:
    A variant of the edge method for the determination of the pre-sampled modulation transfer function (MTF) of digital X-ray imaging devices has been developed and accepted as the standard method in the novel DQE measurement standard IEC 62220-1. An opaque tungsten edge-test device accomplishes the ideal step-like profile of the incident X rays. The edge spread function is measured over a large region across the edge transition that enables an accurate MTF measurement including the 'low-frequency drop'. The method has been applied to different state-of-the-art X-ray imaging detectors, a computed radiography, a CsI-based indirect and an Se-based direct flat-panel detector. The MTF measurement results will be presented. In contrast to the opaque edge device, the commonly used semi-transparent edge-test devices produce scatter radiation that deteriorates the incident X-ray profile, which leads to a systematic overestimation of the MTF.
    [Abstract] [Full Text] [Related] [New Search]