These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A capillary electrophoresis chip with hydrodynamic sample injection for measurements from a continuous sample flow. Author: Büttgenbach S, Wilke R. Journal: Anal Bioanal Chem; 2005 Nov; 383(5):733-7. PubMed ID: 15965683. Abstract: A microchip-based capillary electrophoresis device supported by a microfluidic network made of poly(dimethylsiloxane), used for measuring target analytes from a continuous sample flow, is presented. The microsystem was fabricated by means of replica molding in combination with standard microfabrication technologies, resulting in microfluidic components and an electrochemical detector. A new hydrodynamic sample injection procedure is introduced, and the maximum number of consecutive measurements that can be made with a poly(dimethylsiloxane) capillary electrophoresis chip with amperometric detection is investigated with respect to reproducibility. The device features a high degree of functional integration, so the benefits associated with miniaturized analysis systems apply to it.[Abstract] [Full Text] [Related] [New Search]