These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Theoretical basis of goldmann applanation tonometry]. Author: Wiegand W, Schroeder B, Hager A. Journal: Klin Monbl Augenheilkd; 2005 Jul; 222(7):552-7. PubMed ID: 16034722. Abstract: For more than 50 years Goldmann applanation tonometry has been the internationally accepted method for measuring intraocular pressure. In Goldmann applanation tonometry, however, some basic physical properties are oversimplified and the method has some flaws and restrictions. This paper is intended to promote the understanding of the methodological basis of Goldmann applanation tonometry and describes the most important factors influencing the measurement of intraocular pressure (e. g., corneal thickness, corneal radius, axial length and corneal morphology). Furthermore, the basic principles of other commonly used tonometer devices will be discussed. New developments are anticipated that will measure the true intraocular pressure more accurately than Goldmann applanation tonometry.[Abstract] [Full Text] [Related] [New Search]