These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Consistency of results from the Vettest 8008 clinical biochemistry analyser.
    Author: Little CJ, Gettinby G, Downs J, Marshall CB.
    Journal: Vet Rec; 1992 May 09; 130(19):419-24. PubMed ID: 1609476.
    Abstract:
    Findings are presented on the precision of a clinical biochemistry analyser designed specifically for veterinary use. Twenty biochemical analytes have been examined in detail for variation within and between runs. The results indicate that the analyser can provide high precision for all the analytes with the possible exception of calcium, and suggest that the instrument can be used with confidence in the practice laboratory to aid diagnosis and to monitor biochemical changes in animals receiving treatment.
    [Abstract] [Full Text] [Related] [New Search]