Search MEDLINE/PubMed
Title: Trace anion determination in concentrated hydrofluoric acid solutions by two-dimensional ion chromatography II. Method performance study with a hydroxide eluent and a low noise suppressor. Author: Vermeiren K. Journal: J Chromatogr A; 2005 Aug 26; 1085(1):66-73. PubMed ID: 16106850. Abstract: The two-dimensional ion exclusion chromatography/ion chromatography (ICE-IC) approach is considered to be the method of choice for the determination of trace anions in concentrated hydrofluoric acid (48-50%, w/w). In order to achieve lower detection limits, this method was for the first time used with electrolytically generated and purified hydroxide eluents in combination with a low noise electrochemical suppressor. Compared to carbonate based eluents, the achieved gain in peak height sensitivity for chloride, sulfate, nitrate and phosphate is a factor of 6, 7, 16 and 13, respectively. The instrumental detection limits, based on the background noise, are 2, 0.2, 0.4 and 1.4 microg/kg HF 50% (w/w) for the same anions. Their method detection limits, calculated according to SEMI, are all within the 6-10microg/kg HF 50% (w/w) range and thus at least 10 times lower than the current Tier C grade requirements. The chromatographic run time could be shortened with some 10 min by the use of a relatively fast high-capacity hydroxide selective anion exchange column.[Abstract] [Full Text] [Related] [New Search]