These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Fluorometric determination of ammonium ion by ion chromatography using postcolumn derivatization with o-phthaldialdehyde. Author: Kuo CT, Wang PY, Wu CH. Journal: J Chromatogr A; 2005 Aug 26; 1085(1):91-7. PubMed ID: 16106853. Abstract: A postcolumn fluorometric derivatization method for the determination of trace amounts of ammonium ion (microg/L level) under matrices with high concentrations of sodium and amino acids has been developed. In this method, ammonium ion was determined by ion chromatography combined with fluorometric detection (IC-FL) in less than 16 min. IC was performed in a high-capacity cation-exchange Dionex IonPac CS16 analytical column (250 mm x 5 mm) under isocratic conditions with 30 mM methanesulfonic acid (MSA) as mobile phase at flow-rate 1.0 mL/min. To remove amino acid interference, the postcolumn derivatization based on the reaction of ammonia with o-phthaldialdehyde (OPA) and sulfite was applied. The excitation and emission wavelengths were 364 and 425 nm, respectively. The effects of pH, reaction temperature and time, OPA-reagent composition and concentration, and sample matrix were studied. The linear range and detection limit of this method were similar to the standard method. The IC-FL method with a postcolumn fluorometric derivatization allows the routine determination of ammonium ion in extreme matrices where the ratios of sodium and amino acids to ammonium are up to 2,800,000:1 and 28,000:1, respectively.[Abstract] [Full Text] [Related] [New Search]