These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Automatic registration of portal images and volumetric CT for patient positioning in radiation therapy. Author: Khamene A, Bloch P, Wein W, Svatos M, Sauer F. Journal: Med Image Anal; 2006 Feb; 10(1):96-112. PubMed ID: 16150629. Abstract: The efficacy of radiation therapy treatment depends on the patient setup accuracy at each daily fraction. A significant problem is reproducing the patient position during treatment planning for every fraction of the treatment process. We propose and evaluate an intensity based automatic registration method using multiple portal images and the pre-treatment CT volume. We perform both geometric and radiometric calibrations to generate high quality digitally reconstructed radiographs (DRRs) that can be compared against portal images acquired right before treatment dose delivery. We use a graphics processing unit (GPU) to generate the DRRs in order to gain computational efficiency. We also perform a comparative study on various similarity measures and optimization procedures. Simple similarity measure such as local normalized correlation (LNC) performs best as long as the radiometric calibration is carefully done. Using the proposed method, we achieved better than 1mm average error in repositioning accuracy for a series of phantom studies using two open field (i.e., 41 cm2) portal images with 90 degrees vergence angle.[Abstract] [Full Text] [Related] [New Search]