  • Title: Rapid decline in acute stimulation thresholds with steroid-eluting active-fixation pacing leads.
    Author: Kistler PM, Kalman JM, Fynn SP, Singarayar S, Roberts-Thomson KC, Lindsay CB, Khong U, Sparks PB, Strathmore N, Mond HG.
    Journal: Pacing Clin Electrophysiol; 2005 Sep; 28(9):903-9. PubMed ID: 16176527.
    Abstract:
    BACKGROUND AND AIM: There is an increasing use of active-fixation leads for cardiac pacing, yet concerns remain regarding initial high stimulation thresholds. The aim was to perform a detailed analysis of pacing parameters at the time of implantation to determine when lead repositioning should be considered. METHODS: We performed a prospective observational study of consecutive new pacemaker implants. Detailed analysis of pacing parameters was collected at 2-minute intervals for 10 minutes, and at day 1 and week 8 following implant. RESULTS: Ninety-four patients underwent implantation of 79 dual-chamber and 15 single-chamber pacemakers using active-fixation leads in both chambers. An initial threshold of >1 V was demonstrated in 45/94 (48%) ventricular leads (mean threshold 1.5 +/- 0.3 V). This declined rapidly to 0.9 +/- 0.3 V at 4 minutes (P < 0.01), 0.7 +/- 0.3 V at 10 minutes (P < 0.01), and 0.6 +/- 0.3 V at day 1 (P < 0.01). At day 1, 43/45 leads were <1 V. There were 79 atrial leads. An initial threshold of >1 V (mean 1.7 +/- 0.6 V) was demonstrated in 41/79 (52%) leads falling significantly to 1.1 +/- 0.5 V at 4 minutes (P < 0.01), 0.9 +/- 0.4 V at 10 minutes (P < 0.01), and 0.6 +/- 0.2 V at day 1 (P < 0.01). At 10 minutes, 32 of 41 leads demonstrated a threshold of <1 V with all leads <1 V at day 1. Thresholds were maintained medium term. CONCLUSIONS: Active-fixation leads are commonly associated with initially high thresholds that fall rapidly. An initial threshold of 2 V should be provisionally accepted and retested at 4 minutes. The majority will have a threshold of <1 V the following day. A failure of a high threshold to decline at 4 minutes requires lead repositioning.
    [Abstract] [Full Text] [Related] [New Search]