These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: High-resolution imager for digital mammography: physical characterization of a prototype sensor.
    Author: Suryanarayanan S, Karellas A, Vedantham S, Onishi SK.
    Journal: Phys Med Biol; 2005 Sep 07; 50(17):3957-69. PubMed ID: 16177523.
    Abstract:
    The physical performance characteristics of a high-resolution sensor module for digital mammography were investigated. The signal response of the imager was measured at various detector entrance air kerma and was found to be linear. The spatial resolution was determined by measuring the presampling modulation transfer function, MTF(f), of the system. The noise power spectra, NPS(f), of the system were estimated using 26 kVp: Mo/Mo, 28 kVp: Mo/Rh and 30 kVp: Rh/Rh, with polymethyl methacrylate (PMMA) 'tissue equivalent material' of thickness 20, 45 and 57 mm for each of three x-ray spectra at detector entrance air kerma in the range between approximately 80.2 and 92.3 microGy. The noise equivalent quanta, NEQ(f), and detective quantum efficiencies, DQE(f), for the various spectral conditions were computed. In addition, dose dependence of NPS(f) and DQE(f) was studied at various detector entrance air kerma ranging from 9.4 to 169.7 microGy. A spatial resolution of about 10 cycles mm(-1) was obtained at the 10% MTF(f) level. A small increase in NEQ(f)was observed under higher energy spectral conditions while the DQE(f) decreased marginally. For a given spectrum, increasing PMMA filtration produced negligible change in DQE(f). The estimated DQE values at zero frequency were in the range between 0.45 and 0.55 under the conditions investigated in this study.
    [Abstract] [Full Text] [Related] [New Search]