These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Design of flyback echo-planar readout gradients for magnetic resonance spectroscopic imaging.
    Author: Cunningham CH, Vigneron DB, Chen AP, Xu D, Nelson SJ, Hurd RE, Kelley DA, Pauly JM.
    Journal: Magn Reson Med; 2005 Nov; 54(5):1286-9. PubMed ID: 16187273.
    Abstract:
    The spatial resolution of conventional magnetic resonance spectroscopic imaging-(MRSI) is typically coarse, mainly due to SNR limitations. The increased signal available with higher field scanners and new array coils now permits higher spatial resolution, but conventional chemical shift imaging (phase encoding) limits the spatial coverage possible in a patient-acceptable acquisition time. The "flyback" echo-planar trajectory is particularly insensitive to errors and provides data that are simple to process. In this study, high-efficiency gradient waveforms for flyback echo-planar MRSI were designed and implemented. Normal volunteer studies at 3 T showed the feasibility of acquiring high spatial resolution with large coverage in a short scan time (2048 voxels in 2.3 min and 4096 voxels in 8.5 min). The trajectories were insensitive to errors in timing and require only a modest (10 to 30%) penalty in SNR relative to conventional phase encoding using the same acquisition time.
    [Abstract] [Full Text] [Related] [New Search]