PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A digital beamformer for high-frequency annular arrays.
    Author: Brown JA, Lockwood GR.
    Journal: IEEE Trans Ultrason Ferroelectr Freq Control; 2005 Aug; 52(8):1262-9. PubMed ID: 16245596.
    Abstract:
    Digital transmit and receive beamformers for a 45-MHz, 7-element annular array are described. The transmit beamformer produces 0- to 80-Vpp monocycle pulses with a timing error of less than +/-125 ps. Up to four adjustable transmit focal zones can be selected. The dynamic receive beamformer uses a variable frequency sampling technique in which the frequency of analog-to-digital conversion on each channel is adjusted as the signals are received. The variable frequency clock signals required to trigger analog-to-digital conversion are obtained using a pair of high-frequency field-programmable gate arrays and a precision quartz oscillator. The gate arrays are also used to sum the digitized signals. A maximum receive beamformer timing error of less than +/-900 ps was obtained on each channel. The performance of the combined transmit and receive beamformer was tested by imaging wire phantoms. Images of CD-1 mice were also generated. The system produced images with a dynamic range of 60 dB.