These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Simultaneous determination of four fluoroquinolone residues in fish by ion-pair high performance liquid chromatography]. Author: Guo G, Pan W, Su D, Chen H. Journal: Se Pu; 2005 Jul; 23(4):401-3. PubMed ID: 16250452. Abstract: An ion-pair high performance liquid chromatographic method with a fluorescence detector was established to determine four fluoroquinolone residues in fish. The chromatographic conditions were as follows: Waters microBondapak C18 column (3.9 mm i.d. x 300 mm, 10 microm) was used at 40 degres C with 11 mmol/L tetrabutylammonium bromide solution (pH 3.0) acetonitrile (94: 6, v/v) as the mobile phase with a flow rate of 1.0 mL/min, at the excitation wavelength of 280 nm and emission wavelength of 460 nm. The detection limit for fish was 1 microg/kg. The linear range was from 6-100 microg/kg, and correlation coefficients were more than 0.9995. At the levels of 10, 50 and 100 microg/kg, the recoveries were 76%-100% with relative standard deviations of less than 7%. The method developed can meet the requirement for residue analysis.[Abstract] [Full Text] [Related] [New Search]