These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Comparison of cardiac troponin I levels in serum and sodium-citrate plasma using the ACCESS 2 immunoassay. Author: Wang YS, Feng CF, Er TK, Tsai LY, Chen CY, Chang SJ, Chen BH. Journal: Ann Clin Lab Sci; 2005; 35(4):453-4. PubMed ID: 16254264. Abstract: The aim of this study was to determine if sodium-citrate plasma is a suitable substitute for serum for cardiac troponin I (cTnI) measurement using the Beckman ACCESS 2 immunoassay. Linear regression analysis of cTnI levels in paired serum and sodium-citrate plasma samples from 80 patients provided a y-intercept of -0.0193 and a slope of 1.044. The correlation was good [r=0.995; 95% confidence interval 1.017 to 1.068, p <0.0001]. The paired results were not significantly different by Student's t-test. Serum and sodium-citrate plasma appear to be interchangeable as samples for cTnI measurement using the ACCESS 2 assay. Furthermore, the use of plasma reduces the turnaround time and may avoid false positive cTnI results due to residual fibrin.[Abstract] [Full Text] [Related] [New Search]