These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A Fourier transform Raman spectroscopy analysis of the degree of conversion of a universal hybrid resin composite cured with light-emitting diode curing units. Author: Lindberg A, Emami N, van Dijken JW. Journal: Swed Dent J; 2005; 29(3):105-12. PubMed ID: 16255354. Abstract: The degree of conversion (DC), of a universal hybrid resin composite cured with LED curing units with low and high power densities and a 510 mW/cm2 quartz tungsten halogen unit, was investigated with Fourier Transform Raman spectroscopy. Three curing depths (0, 2, 4mm) and 0 and 7 mm light guide tip - resin composite (LT - RC) distances were tested. The DC of the LED units varied between 52.3% - 59.8% at the top surface and 46.4% - 57.0% at 4 mm depth. The DC of specimen cured with a 0 mm LT- RC distance at 4 mm depth varied between 50.8% - 57.0% and with 7 mm distance between 46.4% - 55.4%. The low power density LED unit showed a significantly lower DC for both distances at all depth levels compared to the other curing units (p < 0.05). Significant differences between the other curing units were only found at the 4 mm depth level cured from 7 mm distance (p < 0.05). The reduction in DC by increasing LT- RC distance was less than 10% for all curing units. It can be concluded that the improved LED curing units could cure the studied resin composite to the same DC as the control unit.[Abstract] [Full Text] [Related] [New Search]