These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Proposed quality specifications for the imprecision and inaccuracy of analytical systems for clinical chemistry. Author: Fraser CG, Petersen PH, Ricos C, Haeckel R. Journal: Eur J Clin Chem Clin Biochem; 1992 May; 30(5):311-7. PubMed ID: 1627730. Abstract: A Working Group of the European Group for the Evaluation of Reagents and Analytical Systems in Laboratory Medicine proposes, after detailed study of the advantages and disadvantages of available strategies, the following quality specifications for analytical systems for clinical chemistry. Total imprecision should be: (a) less than one-half of the average within-subject biological variation, or (b) less than the state of the art achieved by the best 0.20 fractile of laboratories, whichever is the less stringent. The second approach may be used when data on biological variation do not exist. Inaccuracy should be: (a) less than one-quarter of the group (within- plus between-subject) biological variation, or (b) less than one-sixteenth of the reference interval, when data on group biological variation do not exist, or (c) less than twice the ideal imprecision, if the above specifications are too demanding.[Abstract] [Full Text] [Related] [New Search]