These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Micro gas analysis system for measurement of atmospheric hydrogen sulfide and sulfur dioxide. Author: Ohira S, Toda K. Journal: Lab Chip; 2005 Dec; 5(12):1374-9. PubMed ID: 16286968. Abstract: A honeycomb structure microchannel scrubber was developed to achieve efficient and stable gas collection. A thin porous membrane was pasted on a microchannel by the adhesive force of a fresh polydimethylsiloxane surface. The microchannel scrubber achieved much more efficient gas collection than conventional impingers and diffusion scrubbers. Two sets of the microchannel scrubbers and detectors were integrated in a 10 cm x 9 cm plastic board to create a micro gas analysis system (microGAS) for simultaneous measurements of H2S and SO2. The whole system including a battery was incorporated in a carrying case 34 cm W x 16 cm D x 17 cm H for use in the field. Liquid flows at 30 microl min(-1) were obtained by bimetal micropumps. The estimated detection limits were 0.1 ppbv for H2S and 1 ppbv for SO2. The system was demonstrated for real atmospheric gas analysis, and the results agreed well with data concurrently obtained by ion chromatography coupled with a cylindrical diffusion scrubber. The system we developed allowed automated continuous analyses in the field and achieved a much higher time resolution compared to those by ion chromatographic analysis.[Abstract] [Full Text] [Related] [New Search]