  • Title: Off-axis cavity ringdown spectroscopy: application to atmospheric nitrate radical detection.
    Author: Ayers JD, Apodaca RL, Simpson WR, Baer DS.
    Journal: Appl Opt; 2005 Nov 20; 44(33):7239-42. PubMed ID: 16318196.
    Abstract:
    Atmospheric nitrate radicals (NO3) are detected using off-axis cavity ringdown spectroscopy (CRDS) for the first time to our knowledge with a room-temperature continuous-wave (cw) diode laser operating near 662 nm. A prototype instrument was constructed that achieved a 1sigma absorption sensitivity of 5 x 10(-10) cm(-1) Hz(-1/2), corresponding to a 1.4 part per trillion by volume 2sigma detection limit in 4.6 s at 80 degrees C. This sensitivity is a significant improvement over a recent implementation of off-axis cavity-enhanced absorption spectroscopy and comparable to that of the most advanced cw CRDS and pulsed CRDS applications for atmospheric detection of NO3. A comparison of measurements of ambient air in Fairbanks, Alaska, recorded with the off-axis CRDS instrument and a previously characterized conventional cw CRDS instrument showed good agreement (R2 = 0.97).