These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Monte Carlo simulation of intrinsic count rate performance of a scintillation camera for diagnostic images. Author: Mowlavi AA, de Denaro M, Fornasier MR, Binesh A. Journal: Appl Radiat Isot; 2006 Mar; 64(3):390-5. PubMed ID: 16343910. Abstract: This paper describes Monte Carlo simulation of intrinsic count rate performance of a scintillation gamma camera. MCNP Monte Carlo code was employed to calculate pulse height spectrum and detector efficiency. A custom code written in Fortran language was then developed to simulate, by Monte Carlo method, the distortion in pulse height spectrum due to the pile-up effect for paralyzable and nonparalyzable systems. The results of the simulations, compared with the experimental measurement of count rate performance, showed a good agreement between the two different approaches.[Abstract] [Full Text] [Related] [New Search]