These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of theophylline concentration in serum by chemiluminescent immunoassay. Author: Zhou MX, Guan CY, Chen G, Xie XY, Wu SH. Journal: J Zhejiang Univ Sci B; 2005 Dec; 6(12):1148-52. PubMed ID: 16358370. Abstract: OBJECTIVE: This study aimed to establish chemiluminescent immunoassay (CLIA) for quantitative determination of theophylline levels in human serum. METHODS: To measure the concentration of theophylline (n=122) and evaluate the assay. RESULTS: The linear range of the CLIA method was 0.51-40 mg/L (Y=1.02X+0.44, r=0.995). The intra and inter CV (coefficient variance) of CLIA were 3.20% and 3.57%, respectively. The average recovery rate was 102.3%. This method was free from interference by brilirubin (<200 micromol/L), hemoglobin (<10 g/L), and triglycerides (<15 mmol/L). CONCLUSION: This method is simple, convenient and precise for clinical pharmacokinetics study of theophylline.[Abstract] [Full Text] [Related] [New Search]