These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Distal embolus protection with an intra-aortic filter during stent-graft repair of a severely atherosclerotic thoracic aortic aneurysm.
    Author: Sanada J, Matsui O, Ohtake H, Kimura K, Kusanagi M, Terayama N.
    Journal: J Endovasc Ther; 2005 Dec; 12(6):642-6. PubMed ID: 16363892.
    Abstract:
    PURPOSE: To assess the technical feasibility and efficacy of distal embolus protection with an intra-aortic filter during stent-graft repair in a patient diagnosed with a shaggy aorta. TECHNIQUE: In a 75-year-old man with a severely atherosclerotic thoracic aortic aneurysm, stent-graft repair was combined with transposition of the arch branches to the ascending aorta under thoracotomy. A filtration-type embolus protection device with a nitinol basket and polyester fabric was introduced through a 12-F sheath and opened in the supraceliac aorta during the stent-graft procedure; it was safely pulled back into the sheath after the stent-grafts were implanted. The stent-graft repair was successful, and abundant atheromatous debris was captured in the filter. Neither procedure-related embolic events nor neurological deficits were observed. CONCLUSIONS: Use of the intra-aortic filter device to prevent distal embolism during thoracic stent-graft repairs may be feasible and efficacious in severely atherosclerotic patients.
    [Abstract] [Full Text] [Related] [New Search]