These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Single-bounce monocapillaries for focusing synchrotron radiation: modeling, measurements and theoretical limits. Author: Huang R, Bilderback DH. Journal: J Synchrotron Radiat; 2006 Jan; 13(Pt 1):74-84. PubMed ID: 16371711. Abstract: Single-bounce hollow glass capillaries with ellipsoidal shapes have been used at the Cornell High Energy Synchrotron Source recently for various microbeam experiments, with focal spot sizes from 12 to 23 microm, divergences from 2 to 8 mrad, intensities up to 450 times the intensities of incident X-rays, and working distances up to 55 mm. Simple formulae are developed in this paper to explain capillary performance given the X-ray source size, capillary dimensions and slope errors. Capillary length is optimized for best focusing performance. Capillary fabrication accuracy is reported and capillary X-ray tests confirm the focusing properties expected from formulae. The application of capillaries to third-generation X-ray sources and future energy-recovery linac X-ray sources are discussed.[Abstract] [Full Text] [Related] [New Search]