These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Use of flow injection multisite detection as a novel approach for blank signal correction in a spectrophotometric determination. Author: Oliveira TA, Mesquita RB, Lima JL, Rangel AO. Journal: J AOAC Int; 2005; 88(5):1511-5. PubMed ID: 16386002. Abstract: A flow injection multisite detection system was developed for correction of the sample blank in a colorimetric determination. By using detector relocation, a single spectrophotometer is used for the sequential reading of the sample blank and of the colored product; the sample crosses the flow cell to measure intrinsic absorption, the color reagent is subsequently added, and the flow cell is relocated to provide the reading of the resulting plug. The subsequent detector relocation immediately after peak maximum increases the determination frequency. This strategy was tested in the colorimetric determination of iron in soil samples based on the reaction of Fe (II) with 1,10-phenanthroline after reduction of Fe (III) to Fe (II) by ascorbic acid. The results obtained for 15 soil extracts were in good agreement with those obtained by the reference procedure. Relative standard deviations better than 4% were obtained, with a sampling rate of 30/h.[Abstract] [Full Text] [Related] [New Search]