These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Clinically relevant differences in prothrombin time and INR values related to blood sample collection in plastic vs glass tubes. Author: Fiebig EW, Etzell JE, Ng VL. Journal: Am J Clin Pathol; 2005 Dec; 124(6):902-9. PubMed ID: 16416740. Abstract: We compared prothrombin times (PTs) and international normalized ratios (INRs) for blood samples drawn into plastic vs glass collection tubes. We collected 60 venous blood samples into 4.5-mL glass and 2 plastic tubes (2.7 and 3.5 mL). An additional 153 samples, including 63 from warfarin-anticoagulated patients, were collected only in glass and 2.7-mL plastic tubes. The PTs and INRs were determined following routine laboratory procedures. A subset of 35 frozen aliquot samples was analyzed with a different instrument-reagent combination. The PTs and INRs for samples in plastic tubes were significantly lower than for samples in glass tubes. The mean INR differences increased with INR magnitude from approximately -0.1 (INR, 1.5) to -0.7 (INR, 4.5). Of the plastic tube INRs, 50% were more than 10% lower than INRs from samples collected in glass tubes. Therapeutic monitoring based on plastic-tube INRs could result in higher doses of warfarin.[Abstract] [Full Text] [Related] [New Search]