These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Comparison of face masks in the bag-mask ventilation of a manikin.
    Author: Redfern D, Rassam S, Stacey MR, Mecklenburgh JS.
    Journal: Eur J Anaesthesiol; 2006 Feb; 23(2):169-72. PubMed ID: 16426473.
    Abstract:
    BACKGROUND AND OBJECTIVE: We conducted a study investigating the effectiveness of four face mask designs in the bag-mask ventilation of a special manikin adapted to simulate a difficult airway. METHODS: Forty-eight anaesthetists volunteered to bag-mask ventilate the manikin for 3 min with four different face masks. The primary outcome of the study was to calculate mean percentage leak from the face masks over 3 min. Anaesthetists were also asked to rate the face masks using a visual analogue score. RESULTS: The single-use scented intersurgical face mask had the lowest mean leak (20%). This was significantly lower than the mean leak from the single-use, cushioned 7,000 series Air Safety Ltd. face mask (24%) and the reusable silicone Laerdal face mask (27%) but not significantly lower than the mean leak from the reusable anatomical intersurgical face mask (23%). CONCLUSIONS: There was a large variation in both performance and satisfaction between anaesthetists with each design. This highlights the importance of having a variety of face masks available for emergency use.
    [Abstract] [Full Text] [Related] [New Search]