These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Vestibular-evoked myogenic potentials in cochlear implant children. Author: Jin Y, Nakamura M, Shinjo Y, Kaga K. Journal: Acta Otolaryngol; 2006 Feb; 126(2):164-9. PubMed ID: 16428194. Abstract: CONCLUSIONS: Our results suggest that the sacculi of most children with cochlear implants can easily be damaged, as shown by the absence of vestibular-evoked myogenic potentials (VEMPs) in response to click stimuli. Also, in most of the children, the vestibular nerve was seemingly not stimulated by the cochlear implant. These results suggest that electrical stimulation at the C level can stimulate the cochlear nerve; however, this stimulation did not spread to the vestibular nerve in our children. In some children with Mondini dysplasia or vestibulocochlear nerve abnormality, the vestibular nerve was stimulated when the cochlear implant device was on, because of a VEMP response to electrical stimulation. OBJECTIVE: To clarify the diagnostic value of VEMPs in cochlear implant patients. MATERIAL AND METHODS: The click-evoked myogenic potentials of 12 children who underwent cochlear implantation surgery were investigated. The latency and amplitude of the VEMP responses were measured. RESULTS: Before surgery, 6 of the 12 children showed normal VEMPs, 1 showed a decrease in the amplitude of VEMPs and five showed no VEMP response. After surgery, with the cochlear implant device off, 1 child showed a decreased VEMP and 11 showed no VEMPs. With the cochlear implant device on, four children showed VEMPs and eight did not.[Abstract] [Full Text] [Related] [New Search]