Title: Simple spectrophotometric flow injection system with an in-valve minicolumn for enhancement during the determination of chromium(III) using EDTA. Author: Lapanantnoppakhun S, Kasuwas S, Ganranoo L, Jakmunee J, Grudpan K. Journal: Anal Sci; 2006 Jan; 22(1):153-5. PubMed ID: 16429793. Abstract: A simple spectrophotometric flow injection (FI) procedure for the determination of Cr(III) using ethylenediaminetetraacetic acid (EDTA) was developed. An FI system with a column packed with Amberlite IR-120(H) was employed for sample pretreatment. This leads to the possibility of a single standard calibration. A linear calibration in a range of 10-27 microg Cr(III) was obtained with a detection limit of 1 microg Cr(III) and RSD of 2% (18 microg Cr(III), n=12). The proposed procedure was applied for determination of Cr(III) in leachate and dietary supplement samples. The results agreed with those obtained by the standard methods.[Abstract] [Full Text] [Related] [New Search]