These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Analytical interference in determination of procalcitonin by PCT-Q (Brahms)].
    Author: Désidéri-Vaillant C, Rouby Y, Cardon N, Vinsonneau U, Laborde JP.
    Journal: Pathol Biol (Paris); 2006 May; 54(5):293-5. PubMed ID: 16488089.
    Abstract:
    Procalcitonin (PCT) is a biological marker of infection. We present the cas of a patient who has presented a high concentration of PCT with PCT-Q test (Brahms). At the same time, the concentration of CRP is remained low, which is no physiological. Then, PCT concentration has been determinated with an automatic system (Kryptor-Brahms) and finded at low than 0.5 microg/l. Brahms company has searched an analytical interference: human anti-mouse antibodies (HAMA) were positive. If PCT concentration stay a marker of infection for the most part, this case show that biologists have to keep in mind that immunological assays remain submitted to interferences.