Search MEDLINE/PubMed
Title: SMART tubing presents an increased risk of disconnection during extracorporeal circulation. Author: Newling R, Morris R. Journal: J Extra Corpor Technol; 2005 Dec; 37(4):400-1. PubMed ID: 16524161. Abstract: A number of products exhibiting biocompatible features have been developed for use in extracorporeal blood circuits during cardiopulmonary bypass procedures. While attention has been focused on biocompatibility features of the blood-circuit interface, a number of issues applicable in clinical use of these circuits have arisen. Surface Modifying Additive Technology (SMART; Cobe Cardiovascular, Arvarda, CO) is one such technology. In this product, the structure of normal polyvinylchloride (PVC) tubing is altered through the blending of two copolymers to give a more biocompatible blood to plastic interface. In this study, we examined the in vitro mechanical ability of random samples (n = 10) of SMART and standard PVC tubing to withstand axial tension when the tubing was placed over a single barb of a connector. The tension required to remove the SMART tubing from the connector (83.3 +/- 7.3 [SD] N), was significantly less than standard PVC tubing (115.6 +/- 15.9 N; p < .0001, unpaired t test). The SMART tubing exhibited a 28% reduction in tubing to connector adhesion, which may have a significant effect on extracorporeal circuit disconnection and overall patient safety.[Abstract] [Full Text] [Related] [New Search]