These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Simultaneous detection of HCV and HIV-1 by SYBR Green real time multiplex RT-PCR technique in plasma samples. Author: Gibellini D, Gardini F, Vitone F, Schiavone P, Furlini G, Re MC. Journal: Mol Cell Probes; 2006; 20(3-4):223-9. PubMed ID: 16537101. Abstract: This paper describes the development of a SYBR Green-based multiplex real time RT-PCR for the simultaneous detection of HCV and HIV-1 genomes in plasma samples. Viral genomes were identified in the same sample by their distinctive melting temperature (Tm) which are 81.6 and 86.5 degrees C for HIV-1 gag 142 bp amplicon and HCV 5'-NCR region 226 bp amplicon, respectively. Analysis of known scalar concentrations of reference plasma indicated that the multiplex procedure detects at least 500 copies/ml of both HIV-1 and HCV. In addition, we also assayed HIV-1 and HCV viral load in 30 co-infected patients and in 15 blood donors, confirming the sensitivity and specificity of the assay. This method may represent a useful alternative method for the detection of HIV-1/HCV co-infection, reliable for a rapid and relatively inexpensive screening of blood donors.[Abstract] [Full Text] [Related] [New Search]