Search MEDLINE/PubMed
Title: Evaluation and comparison of two immunoturbidimetric assays for the heterologous determination of porcine serum C-reactive protein. Author: Kjelgaard-Hansen M, Martínez-Subiela S, Petersen HH, Jensen AL, Ceron JJ. Journal: Vet J; 2007 May; 173(3):571-7. PubMed ID: 16546421. Abstract: The objective was first to evaluate the performance (intra- and inter-assay imprecision, inaccuracy, detection limit and overlap performance) of two different commercially available human C-reactive protein (CRP) immunoturbidimetric assays (TIA) applied to porcine serum CRP; and secondly to investigate inter-laboratory discrepancies and the influence of calibration materials when comparing results. The intra-laboratory performance seemed acceptable for clinical purposes for both assays with excellent overlap performance, acceptable detection limits, no significant inaccuracies and low imprecision. Although a significant discrepancy in the measured CRP levels was observed when comparing results, the methods agreed significantly on diagnostic classification, suggesting a low clinical significance. Both assays performed reliably for the heterologous determination of porcine serum CRP for clinical purposes. The results emphasise the need to establish species-specific standards to decrease inter-laboratory discrepancies.[Abstract] [Full Text] [Related] [New Search]