These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Profilometric and standard error of the mean analysis of rough implant surfaces treated with different instrumentations.
    Author: Ramaglia L, di Lauro AE, Morgese F, Squillace A.
    Journal: Implant Dent; 2006 Mar; 15(1):77-82. PubMed ID: 16569965.
    Abstract:
    PURPOSE: This study evaluated, in vitro, the effects of different instrumentations used in the treatment of peri-implantitis on implant surfaces coated with hydroxyapatite or titanium plasma spray (TPS). MATERIALS AND METHODS: There were 14 cylindrical rough implants used, including 7 hydroxyapatite and 7 TPS coated. Split in 2 parts for a total of 24 experimental surfaces, implants were treated with a stainless-steel curette, plastic curette, ultrasonic scaler tip, and air-powder-water spray. There was 1 hydroxyapatite and 1 TPS implant used as controls. Profilometry and scanning electron microscopy were used to examine instrumented surfaces for variations in surface topography. RESULTS: All experimental procedures determined changes on tested rough implant surfaces. Such alterations were related to the implant coating material, and the procedure consisting in coating removal and/or leveling of surface roughness. CONCLUSION: Although a plastic curette and air-powder-water spray induced less implant surface alterations, these instrumentations left deposits on the surface that may affect, in vivo, the tissue healing process.
    [Abstract] [Full Text] [Related] [New Search]