Title: Profilometric and standard error of the mean analysis of rough implant surfaces treated with different instrumentations. Author: Ramaglia L, di Lauro AE, Morgese F, Squillace A. Journal: Implant Dent; 2006 Mar; 15(1):77-82. PubMed ID: 16569965. Abstract: PURPOSE: This study evaluated, in vitro, the effects of different instrumentations used in the treatment of peri-implantitis on implant surfaces coated with hydroxyapatite or titanium plasma spray (TPS). MATERIALS AND METHODS: There were 14 cylindrical rough implants used, including 7 hydroxyapatite and 7 TPS coated. Split in 2 parts for a total of 24 experimental surfaces, implants were treated with a stainless-steel curette, plastic curette, ultrasonic scaler tip, and air-powder-water spray. There was 1 hydroxyapatite and 1 TPS implant used as controls. Profilometry and scanning electron microscopy were used to examine instrumented surfaces for variations in surface topography. RESULTS: All experimental procedures determined changes on tested rough implant surfaces. Such alterations were related to the implant coating material, and the procedure consisting in coating removal and/or leveling of surface roughness. CONCLUSION: Although a plastic curette and air-powder-water spray induced less implant surface alterations, these instrumentations left deposits on the surface that may affect, in vivo, the tissue healing process.[Abstract] [Full Text] [Related] [New Search]