These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Silver(I)-selective electrode based on [Bz2Oxo4(18)dieneS4] tetrathia macrocyclic carrier.
    Author: Singh AK, Saxena P.
    Journal: Anal Bioanal Chem; 2006 May; 385(1):90-5. PubMed ID: 16601953.
    Abstract:
    A polystyrene-based membrane of 7,8:16,17-dibenzo-6,9,15,18-tetraoxo-1,5,10,14-tetrathiacyclooctadeca-7,16-diene [Bz2Oxo4(18)dieneS4] was fabricated using sodium tetraphenylborate (NaTPB) and dioctyl phthalate (DOP) as anion excluder and plasticizing agent. The best performance was obtained from the membrane with the composition ionophore [Bz2Oxo4(18)dieneS4]:polystyrene:DOP:NaTPB, 5:100:150:10 (w/w). The response of the electrode was linear over a wide range of concentration, 1.26 x 10(-6)-1.00 x 10(-1) mol L(-1) for silver ion with a Nernstian slope of 58.4 +/- 0.1 mV per decade and a detection limit of 1.0 x 10(-6) mol L(-1). The electrode was found to be chemically inert and of adequate stability with a response time of 10 s and could be used for a period of 3 months without change of potential. It worked satisfactorily in mixtures containing up to 35% (v/v) non-aqueous content. The proposed membrane sensor had good selectivity for Ag+ over a wide variety of metal ions in the pH range 2.2-8.5. It was successfully used as an indicator electrode in potentiometric titration of silver ion. The electrode was also useful for determination of Ag+ in waste from photographic films.
    [Abstract] [Full Text] [Related] [New Search]