These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: International collaborative study for the calibration of a proposed international standard for thromboplastin, rabbit, plain. Author: Chantarangkul V, van den Besselaar AM, Witteveen E, Tripodi A. Journal: J Thromb Haemost; 2006 Jun; 4(6):1339-45. PubMed ID: 16706980. Abstract: BACKGROUND: A preparation of rabbit brain thromboplastin, provisionally coded 04/162, is proposed as a candidate for the World Health Organization (WHO) International Standard (IS) for thromboplastin (rabbit, plain), meant to replace the IS coded RBT/90 (rabbit, plain), stocks of which are now exhausted. RESULTS: The preparation was calibrated in an international collaborative study involving 21 laboratories from 13 countries and the calibration was performed against the existing WHO-IS (i.e. rTF/95 and OBT/79) and other Certified Reference Materials from the Institute for Reference Materials and Measurements of the European Commission (i.e. CRM149 S) and from the European Action on Anticoagulation (i.e. EUTHR-01). An additional candidate rabbit brain thromboplastin coded as 04/106 was also included in the study. On the basis of predefined criteria (the within- and between-laboratory precision of the calibration and the conformity to the calibration model), 04/162 was the preferred candidate. CONCLUSIONS: The assigned International Sensitivity Index value was 1.15 and the inter-laboratory SD and coefficient of variation were 0.057% and 4.9%, respectively.[Abstract] [Full Text] [Related] [New Search]