These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Neutron dosimetry with 6LiF-rich TL sheet.
    Author: Konnai A, Odano N, Nariyama N, Ohnishi S, Nakajima N, Yamamoto K, Kishi T, Ozasa N, Ishikawa Y.
    Journal: Radiat Prot Dosimetry; 2006; 120(1-4):133-5. PubMed ID: 16709707.
    Abstract:
    A (6)LiF-rich thermoluminescent sheet-type dosemeter ((6)LiF-rich NTL sheet) was developed for neutron 2D dosimetry. The dosemeter utilises the (6)Li(n, alpha)(3)H reaction to detect thermal neutrons. Responses of the (6)LiF-rich NTL sheet to neutrons were measured at the neutron beam irradiation facility for BNCT in JRR-4 Research Reactor at the Japan Atomic Energy Research Institute. Placement of a multi-leaf collimator at the output port of the neutron (beam) irradiation facility, produced either stripe- or round-shaped neutron distributions; the spatial distribution was measured using the developed NTL sheets. Direct measurements of neutron attenuation in water were also carried out using the developed NTL sheet, submersed in a water phantom. In each experiment, NTL sheets having natural abundance (7.9%) of LiF, and (6)LiF-enriched NTL (18.94%) sheet were irradiated under the same conditions. The ratio of thermoluminescence intensities of the (6)LiF-rich NTL sheet to that of the normal NTL sheet was compared to a theoretically calculated value. The experimental measurements are shown to be in good agreement with the calculations.
    [Abstract] [Full Text] [Related] [New Search]