These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: High-throughput quantification of perindopril in human plasma by liquid chromatography/tandem mass spectrometry: application to a bioequivalence study.
    Author: Nirogi RV, Kandikere VN, Shukla M, Mudigonda K, Maurya S, Komarneni P.
    Journal: Rapid Commun Mass Spectrom; 2006; 20(12):1864-70. PubMed ID: 16715478.
    Abstract:
    A simple, sensitive and rapid high-performance liquid chromatography/positive ion electrospray tandem mass spectrometry method was developed and validated for the quantification of perindopril in human plasma. Following liquid-liquid extraction, the analytes were separated using an isocratic mobile phase on a reversed-phase column and analyzed by mass spectrometry in the multiple reaction monitoring mode using the respective [M+H](+) ions, m/z 369/172 for perindopril and m/z 417/234 for the internal standard. The method exhibited a linear dynamic range of 0.1-100 ng/mL for perindopril in human plasma. The lower limit of quantification was 0.1 ng/mL with a relative standard deviation of less than 6.1%. Acceptable precision and accuracy were obtained for concentrations over the standard curve range. A run time of 2.0 min for each sample made it possible to analyze more than 450 human plasma samples per day. The validated method has been successfully used to analyze human plasma samples for application in pharmacokinetic, bioavailability and bioequivalence studies.
    [Abstract] [Full Text] [Related] [New Search]