Title: A practical approach to eliminate autocorrelation artefacts for volume-rate spectral domain optical coherence tomography. Author: Wang RK, Ma Z. Journal: Phys Med Biol; 2006 Jun 21; 51(12):3231-9. PubMed ID: 16757873. Abstract: A simple method is introduced to eliminate the autocorrelation artefacts in ultrafast spectral domain optical coherence tomography (SOCT) by use of the ensemble average of spectra within an individual B scan as the background signal, and then subtracting this from all the A scans within that B scan before performing the FFTs. This is updated continuously frame by frame. The method is tested on a volume-rate (C-mode) SOCT system to image the human fingertip in vivo with a volume rate at 8 s.[Abstract] [Full Text] [Related] [New Search]