  • Title: Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis: the detect supraventricular tachycardia study.
    Author: Friedman PA, McClelland RL, Bamlet WR, Acosta H, Kessler D, Munger TM, Kavesh NG, Wood M, Daoud E, Massumi A, Schuger C, Shorofsky S, Wilkoff B, Glikson M.
    Journal: Circulation; 2006 Jun 27; 113(25):2871-9. PubMed ID: 16769912.
    Abstract:
    BACKGROUND: Delivery of inappropriate shocks caused by misdetection of supraventricular tachycardia (SVT) remains a substantial complication of implanted cardioverter/defibrillator (ICD) therapy. Whether use of optimally programmed dual-chamber ICDs lowers this risk compared with that in single-chamber ICDs is not clear. METHODS AND RESULTS: Subjects with a clinical indication for ICD (n=400) at 27 participating centers received dual-chamber ICDs and were randomly assigned to strictly defined optimal single- or dual-chamber detection in a single-blind manner. Programming minimized ventricular pacing. The primary end point was the proportion of SVT episodes inappropriately detected from the time of programming until crossover or end of study. On a per-episode basis, 42% of the episodes in the single-chamber arm and 69% of the episodes in the dual-chamber arm were due to SVT. Mortality (3.5% in both groups) and early study withdrawal (14% single-chamber, 11% dual-chamber) were similar in both groups. The rate of inappropriate detection of SVT was 39.5% in the single-chamber detection arm compared with 30.9% in the dual-chamber arm. The odds of inappropriate detection were decreased by almost half with the use of the dual-chamber detection enhancements (odds ratio, 0.53; 95% confidence interval, 0.30 to 0.94; P=0.03). CONCLUSIONS: Dual-chamber ICDs, programmed to optimize detection enhancements and to minimize ventricular pacing, significantly decrease inappropriate detection.
    [Abstract] [Full Text] [Related] [New Search]