These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Late stent-graft migration secondary to separation of the uncovered segment from the main body of a Zenith endoluminal graft.
    Author: Ghanim K, Mwipatayi BP, Abbas M, Sieunarine K.
    Journal: J Endovasc Ther; 2006 Jun; 13(3):346-9. PubMed ID: 16784322.
    Abstract:
    PURPOSE: To report a case of type I endoleak secondary to complete disruption of the sutures uniting the uncovered and covered segments of a bifurcated Zenith endoluminal graft, causing displacement and distal migration of the graft main body. CASE REPORT: A 76-year-old man had successful exclusion of an abdominal aortic aneurysm with a Zenith endoluminal graft in 1999. He continued to do well until the 4-year surveillance imaging [computed tomography (CT) and plain abdominal radiography] showed device migration and proximal endoleak, with consequent expansion of the aneurysm. A proximal extension stent-graft was inserted with good seal. The 1-month follow-up CT angiogram showed reduced aneurysm size and no evidence of any leak. CONCLUSION: This case shows that the failure of an endoluminal graft occurs at weak points in the construction of the graft, reinforcing the need for long-term surveillance. If detected promptly, such events can often be treated by another endovascular procedure.
    [Abstract] [Full Text] [Related] [New Search]